No stress rentals! We help you to choose the equipment that you need, we expedite shipment, and follow up.
Input specifications | |
Voltage | 115/130 VAC ± 10%, user selectable |
Frequency | 50/60 Hz ± 5% |
Fuse | 3.15 A, fast acting 250 VAC |
Dielectric Withstand Test Mode | |
Output Rating | 5000 V at 100 mA AC |
Maximum Limit, AC | Range: 0.00 - 99.99 mA Resolution: 0.01 mA Accuracy: ± (2% of setting + 6 counts) |
Minimum Limit, AC | Range: 0.000 - 9.999 mA Resolution: 0.001 mA Accuracy: ± (2% of setting + 6 counts) |
Arc Detection | Range: 0 - 9, 0 disabled |
Ground Fault Interrupt | GFI Trip Current: 450 μA max (AC or DC) HV Shut Down Speed: < 1ms |
Current Display, AC auto range | Range 1: 0.000 - 3.500 mA Range 2: 3.00 - 99.99 mA |
Discharge Time | ≤ 200 ms |
Maximum capacitive load | 0.20 μF < 1 kV 0.10 μF < 2 kV 0.06 μF < 3 kV 0.050 μF < 4 kV 0.040 μF < 5 kV 0.015 μF < 6 kV |
AC Voltage Waveform | Sine Wave, Crest Factor = 1.3 - 1.5 |
Output Frequency | Range: 50 or 60 Hz, User Selectable |
Output Voltage Regulation | ± (1% of output + 5 V) from no load to full load and over input voltage range |
Dwell Timer | Range: AC 0, 0.3 - 999.9 sec (0 = Constant) Accuracy: ± (0.1% of reading + 0.05 sec) |
Ramp Timer | Range: Ramp-Up: 0.1 - 999.9 sec Ramp-Down: AC 0.0 - 999.9 sec Accuracy: ± (0.1% of reading + 0.05 sec) |
Output Short Circuit Current | > 200 mA |
General Specifications | |
Memory | 10 Memories, 3 steps per memory |
Interface | RS-232 Interface standard for entry-level automation |
Mechanical | Bench or rack Mount with tilt up feet |
Dimensions | (W x H x D) 8.46 x 3.5 x 14.57 in., (215 x 89 x 370 mm) |
Weight | 20.96 lbs (9.53 kgs) |
For detailed specifications please see: Associated Research 3780 Data Sheet |