No stress rentals! We help you to choose the equipment that you need, we expedite shipment, and follow up.
Frequency range, RF path A | SMW-B103: 100 kHz to 3 GHz SMW-B106: 100 kHz to 6 GHz SMW-B107: 100 kHz to 7.5 GHz SMW-B112: 100 kHz to 12.75 GHz SMW-B120: 100 kHz to 20 GHz SMW-B131: 100 kHz to 31.8 GHz SMW-B140/-B140N: 100 kHz to 40 GHz |
---|---|
Frequency range, RF path B (optional) see SMW200A data sheet for possible RF path configurations |
SMW-B203: 100 kHz to 3 GHz SMW-B206: 100 kHz to 6 GHz SMW-B207: 100 kHz to 7.5 GHz SMW-B212: 100 kHz to 12.75 GHz SMW-B220: 100 kHz to 20 GHz |
Setting time | SCPI mode: < 1.2 ms, 600 μs typical |
Specified level range | 3 MHz ≤ f < 20 GHz (–120 dBm to 18 dBm PEP) SMW-B131, SMW-B140/-B140N: –120 dBm to +15/+18 dBm (PEP), depending on RF |
---|---|
Setting time | SCPI mode: < 1 ms, 600 μs (typical) SMW-B120/-B131/-B140/-B140N/-B220, with switching of the mechanical step attenuator: < 25 ms |
Harmonics | Level < 10 dBm, CW < -30 dBc SMW-B120/-B131/-B140/-B140N/-B220, f >3.5 GHz < -55 dBc |
---|---|
Nonharmonics | CW or vector modulation with full-scale DC input, level > -10 dBm, carrier offset > 10 kHz, 200 MHz < f ≤ 1500 MHz Standard: < -80 dBc With SMW-B22 option: < -90 dBc |
SSB phase noise | CW, carrier offset = 20 kHz, f = 1 GHz Standard: < -131 dBc, -135 dBc with SMW-B22 option: < -137 dBc, -141 dBc CW, carrier offset = 20 kHz, f = 10 GHz Standard: < -111 dBc, -115 dBc with SMW-B22 option: < -117 dBc, -120 dBc |
More information... | For complete specifications see: Rohde & Schwarz SMW200A Data Sheet |