  • Features & specifications

  • Output peak envelope power from -120dBm to +18dBm
  • Output power -145dBm to +18dBm
  • Frequency resolution 0.01Hz
  • SSB phase noise 20kHz offset: -131dBc/Hz at 1GHz;(typ.-135dBc/Hz)
  • Standard modulations: AM and Pulse

Specifications
Frequency
Frequency range SMA-B103/-B103L
SMA-B106/-B106L
9 kHz to 3 GHz
9 kHz to 6 GHz
Resolution of setting 0.01 Hz
Resolution of synthesis fundamental frequency range 750MHz to 1500MHz:
standard
with SMA-B22 option

5 μHz (nom.)
0.2 μHz (nom.)
Setting time to within < 1 x 10–7 for f > 6.6 MHz or < 35Hz for f < 6.6 MHz
no relay switchover, PLL bandwdith normal
after IEC/IEEE bus delimiter
in ALC OFF S&H mode
after trigger pulse in List mode or Fast Hopping mode
< 1.2 ms
< 5 ms
< 450 μs
Resolution of phase offset setting adjustable in 0.1° steps
Main PLL bandwith settings
with SMA-B22 option
f ≤ 3 GHz
f > 3 GHz
normal, narrow
normal
Multiplier for phase-continuous frequency setting f ≤ 6.6 MHz
6.6 MHz < f ≤ 11.71875 MHz
11.71875 MHz < f ≤ 23.4375 MHz
23.4375 MHz < f ≤ 46.875 MHz
46.875 MHz < f ≤ 93.75 MHz
93.75 MHz < f ≤ 187.5 MHz
187.5 MHz < f ≤ 375 MHz
375 MHz < f ≤ 750 MHz
750 MHz < f ≤ 1500 MHz
1500 MHz < f ≤ 3 GHz
f > 3 GHz
rm = 1/2
rm = 1/128
rm = 1/64
rm = 1/32
rm = 1/16
rm = 1/8
rm = 1/4
rm = 1/2
rm = 1
rm = 2
rm = 4
Phase-continuous frequency setting range
with SMA-B22 option, narrow mode
with SMA-B22 option, wide mode
rm x 1 MHz (nom.)
rm x 5 MHz
rm x 20 MHz
Max. phase-continuous frequency step
with SMA-B22 option
rm x 2 kHz (nom.)
rm x 100 kHz (nom.)
Environmental conditions
Temperature range Operating
Storage
0 °C to +55 °C
-40 °C to +71 °C
Dimensions and weight
Dimensions 465 mm x 108 mm x 495 mm
(18.31 in x 4.25 in x 19.49 in)
Weight when fully equipped 10 kg (22 lb)
Calibration interval