No stress rentals! We help you to choose the equipment that you need, we expedite shipment, and follow up.
Input specifications | |
Voltage | 115/130 VAC ± 10%, Single Phase, User Selection |
Frequency | 50/60 Hz ± 5% |
Fuse | 6.3 A, 250 V Slow Blow |
Dielectric Withstand Test Mode | |
Output Rating | 20 kV at 5 mA DC |
Maximum Limit | Range 1: 0.0 - 999.9 μA Range 2: 1000 - 5000 μA Resolution: 0.1 μA/step Accuracy: ± (2% of setting + 2 counts) |
Voltage Display | Range: 0.00 - 20.00 kV Full scale Accuracy: ± (2% of reading + 20 V) |
Current Display | Range 1: 0.0 - 350.0 μA Range 2: 300 μA - 5000 μA |
DC Output Ripple | ≤ 5% Ripple at 20 kV at 4999 μA, Resistive Load |
Output Regulation | ± (1% of output + 10 V) from no load to full load |
Discharge Timer | No load < 500 ms |
Dwell Timer | Range: DC 0, 0.4 - 999.9 sec (0 = Continuous) |
Ramp Timer | Range: 0.4 - 999.9 sec |
Ground Fault | HV Shut Down Speed < 1ms |
Interrupt | GFI trip current 1 mA max |
General Specifications | |
Memory | 50 Memories w/8 steps per memory |
Interface | Standard USB/RS-232. Optional GPIB |
Mechanical | Tilt up feet |
Dimensions | (W x H x D) 16.93 x 5.24 x 15.75 in., (430 x 133 x 400 mm) |
Weight | 48.9 lbs (22 kgs) |
Shipping weight | 60 lbs (27.22 kgs) |
For detailed specifications please see: Associated Research 7720 Data Sheet |