No stress rentals! We help you to choose the equipment that you need, we expedite shipment, and follow up.
Frequency range | 9 kHz to 3 GHz |
Measurement speed | |
Local measurement and display update rate | 33 ms, (30/s) |
Remote measurement and GPIB transfer rate | 33 ms, (30/s) |
Marker peak search | 300 ms |
Center frequency tune and transfer (RF) | < 90 ms |
Center frequency tune and transfer (μW) | 350 ms |
Offset from carrier signal. Typical in italic. | |
10 kHz | –98 dBc/Hz, –101 dBc/Hz, (Option 1D5) |
100 kHz | –118 dBc/Hz, –122 dBc/Hz |
1 MHz | –125 dBc/Hz, –127 dBc/Hz –133 dBc/Hz, –136 dBc/Hz (with option 120) |
10 MHz | –131 dBc/Hz, –136 dBc/Hz –137 dBc/Hz, –141 dBc/Hz (with option 120) |
Mechanical input attenuator range | 0 to 75 dB in 5 dB steps |
Average continuous power | +30 dBm (1 W) |
Peak pulse power | +50 dBm (100 W) |
DC voltage | DC coupled: 0 Vdc (option UKB) AC coupled 100 Vdc 50 Vdc (Option UKB) |
Displayed average noise level (dBm) (input terminated, 0 dB attenuation, sample detector) specifications Typical values shown in italic |
|
Frequency: 1 to 10 MHz | -139 dBm |
Frequency: 10 to 500 MHz 500 MHz to 1 GHz |
-136 dBm, -140 dBm |
Frequency: 1 to 1.5 GHz 1.5 to 2 GHz |
-135 dBm, -140 dBm |
Frequency: 2 to 3 GHz | -133 dBm, -140 dBm |
Displayed average noise level (dBm) with RF preamplifier Typical values shown in italics |
|
Frequency: 1 to 10 MHz | -152 dBm |
Frequency: 10 MHz to 1 GHz 1 to 2 GHz |
-152 dBm, -156 dBm |
Frequency: 2 to 3 GHz | -151 dBm, -154 dBm |
Resolution bandwidth switching uncertainty | |
1 Hz, 3 Hz RBW | ± 0.3 dB (Option 1DR, Option 1D5) |
10 Hz, 30 Hz RBW | ± 0.3 dB (Option 1DR) |
100 Hz, 300 Hz RBW | ± 0.3 dB (Option 1DR) |
1 kHz to 1.5 MHz RBW 1.5 to 3 MHz RBW |
± 0.3 dB |
5 MHz RBW | ± 0.6 dB |
Operating temperature | 0 to +55 °C |
Storage temperature | -40 to +75 °C |
EMI compatibility | Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class A. Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class Ba (Option 060) |
Military specifications | Type tested to the environmental specifi cations of MIL-PRF-28800F Class 3 |
AC operation | 195 to 250 Vrms, 47 to 66 Hz Power consumption < 300 W |
DC operation | 12 to 20 Vdc, < 200 W power consumption |
Internal data storage | 200 traces or states/8.0 MB |
Dimensions | Width to outside of instrument handle: 416 mm (16.4 in) Overall height: 222 mm (8.75 in) Depth from front frame to rear frame: 409 mm (16.1 in) |
Weight | Instrument: 15.5 kg (34.2 lbs) Shipping: 27.4 kg (60.4 lbs) |
For detailed specifications please see: Keysight E4402B Data Sheet |