| Measurement line type | Single-phase 2-wire, single-phase 3-wire, three-phase 3-wire, and three-phase 4-wire systems (50/60 Hz) | 
|---|---|
| Number of systems that can be measured (for systems that share the same voltage) | 1P2W - 4 systems 1P3W - 2 systems 3P3W2M - 2 systems 3P3W3M - 1 system 3P4W - 1 system 3P4W4I - 1 system  | 
		
| Parameters | Voltage, current, active power, reactive power, apparent power, power factor, integrated value, frequency, and harmonics. | 
| Measurement method | Simultaneous digital sampling of voltage and current, PLL synchronization or a fixed clock (50/60 Hz) | 
| Input methods | Isolated voltage input and isolated current input using a clamp-on sensor | 
| Display | 5.7 in. backlit LCD (320 x 240 dots.) | 
| Input resistance | Voltage: 2.0 MΩ ± 10% (differential input) Current: 200kΩ ± 10%  | 
		
| Maximum input | Voltage input: 780 Vrms AC, peak value: 1103V Current input: 1.7 Vrms AC, peak value: 2.4 V  | 
		
| Measurement method (Voltage & Current)  | 
			True RMS method | 
|---|---|
| Measurement display (Voltage & Current)  | 
			Measurement of 3 voltage lines and 3 or 4 current lines is possible when using 3-phase 3 wire and 3-phase 4-wire systems. | 
| Measurement display (Active power)  | 
			For three-phase 3-wire (the 3P3W3M setting), refer to the display for phase power values | 
| Polarity display (Active power)  | 
			For consumption: no signal for regeneration | 
| Reactive Power measurement | ON: Measures the reactive power directly using the reactive power measurement method OFF: Calculates the reactive power from the measured values of voltage, current, and active power  | 
		
| Polarity display | For lag phase (LAG: current is slower than voltage) For lead phase (LEAD: current is faster than voltage) (Reactive power measurement: ON)  | 
		
| Polatity display Apparent power  | 
			No polarity | 
| Measurement range (Power factor measurement)  | 
			-1.0000 (lead) to 0.0000 to + 1.0000 (lag) | 
| Polarity display (Power factor measurement)  | 
			No polarity | 
| Measurement range (Frequency)  | 
			40.000 to 70.000 Hz | 
| Input area for guaranteed accuracy | Within 10 to 110% of the range (for sine wave input) | 
| Operating environment | Indoors, up to 2000m above sea level | 
|---|---|
| Operating temperature | 0 to 40°C, 80% RH or less (non-condensing) | 
| Storage temperature | -10 to 50°C, 80% RH or less (non-condensing) | 
| Withstand voltage (50/60 Hz for 15 sec.)  | 
			5.55 kVrms AC between the voltage input terminal and the instrument casing 3.32 kVrms AC between the voltage input terminal and the current input terminal/external interface terminal 2.3 kVrms AC between the Power Supply and the instrument casing 1.39 kVrms AC between the power supply and the current input terminal/external input terminal  | 
		
| Power supply voltage rating | 100 to 240 V AC, 50/60 Hz | 
| Maximum rated power | 30 VA | 
| Dimensions & weight | (W x H x D), (210 x 160 x 60)mm, 1.2 kg | 
| Standard accessories | Quick Start Guide Documentation CD Manual for PQA-HiViewpro PQA-HiView Pro CD Battery Pack Power Cord Hard Carrying Case Meter LAN Crossover Cable Model 9661: 500 amp CTs, qty. 4 4 Leads: (Yellow, Red, Blue, and Black) Compact Flash Adapter 32 MB (or greater) Compact Flash Memory Card 12 Volt AC Adapter 400Hz Measurement Function Manual Mode Instruction Manual Multi Card Reader with USB Cable  |